Operation
Instrument Controls Menu
3-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
PUMP:
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Set Heater menu allows the user to set the status of the heater, RH
threshold, and temperature threshold in various modes.
In the Main Menu, choose Instrument Controls > Set Heater.
HEATER:
>CONTROL OFF
RH THRESHOLD 40 %
TEMP THRESHOLD 30
o
C
RANGE AVG DIAGS ALARM
The Control screen is used to set the status of the heater to either
ON/OFF, RH threshold, or TEMP threshold.
In the Main Menu, choose Instrument Controls > Set Heater >
Control.
HEATER CONTROL:
CURRENTLY: OFF
SET TO: RH ?
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The RH Threshold screen is used to change the RH threshold value in
percent. The RH threshold is typically compatible with the gravimetric
method being used for comparison.
In the Main Menu, choose Instrument Controls > Set Heater > RH
Threshold.
Set Heater
Control
RH Threshold