Model 5030i SHARPInstruction Manual Synchronized Hybrid Ambient Real-time Particulate Monitor Part Numbe

Strany 2

About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific

Strany 3 - WEEE Compliance

Operation Volumetric Conditions 3-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INSTRUMENT ID: CURRENTLY: 14

Strany 4

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-39 Choices are Concentrations, Other Measurements

Strany 5

Operation Volumetric Conditions 3-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CHOOSE STREAM DATA: >CONCENTRATIONS OTHER

Strany 6 - Damage Alerts

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-41 OTHER MEASUREMENTS: >NONE BARO PRES VAC

Strany 7

Operation Volumetric Conditions 3-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Control

Strany 8 - FCC Compliance

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-43 The Netmask screen is used to edit the netmas

Strany 9 - EEE Symbol

Operation Volumetric Conditions 3-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific HOST NAME: CURRENT: ISERIES IS

Strany 10 - Where to Get Help

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-45 Note The digital outputs may take up to one s

Strany 11 - Contents

Operation Volumetric Conditions 3-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Alarms screen allows the user to select a

Strany 12

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-47 Not all of the I/O available in the instrument

Strany 13

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual vii Contents Introduction ...

Strany 14

Operation Volumetric Conditions 3-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CHOOSE ACTION: >NONE AOU

Strany 15

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-49 ALLOW OVER/UNDER RANGE: CURRENTLY:

Strany 16

Operation Volumetric Conditions 3-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific analog output channel. See Table 3–4 for a lis

Strany 17

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-51 Output Zero (0)% Value Full-Scale 100% Value

Strany 18

Operation Volumetric Conditions 3-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Concentrations Other Measurements Analog Inp

Strany 19

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-53 ● In the Main Menu, choose Instrument Control

Strany 20

Operation Volumetric Conditions 3-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANALOG INPUT UNITS: CURRENTLY: V

Strany 21

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-55 The Table Point submenu allows the user to se

Strany 22

Operation Volumetric Conditions 3-56 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific TABLE POINT 01 USER VAL: CURRENTLY:

Strany 23 - Figures

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-57 Note The service mode should be turned off wh

Strany 24

Contents viii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Firmware Overview ...

Strany 25

Operation Diagnostics Menu 3-58 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Controls >

Strany 26

Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-59 PROGRAM VERSIONS: PRODUCT: MODEL 5030i VE

Strany 27

Operation Diagnostics Menu 3-60 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INTERFACE BOARD VOLTAGES: 3.3 SUPPLY 3.3

Strany 28

Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-61 NEPH BOARD VOLTAGES: 3.3 SUPPLY 3.3 V

Strany 29 - Introduction

Operation Diagnostics Menu 3-62 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific FLOW: 16.67 LPM

Strany 30 - Principle of Operation

Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-63 CRn 0.00 Bq/m3

Strany 31 - Principle of

Operation Diagnostics Menu 3-64 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DIGITAL INPUTS: >INPUT 1 1 INPU

Strany 32

Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-65 operation upon exiting this screen. The following e

Strany 33

Operation Alarms Menu 3-66 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The alarms menu allows the user to view a set of alarms

Strany 34

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-67 INSTRUMENT ALARMS: >FILTER TAPE COUNTER OK FILTE

Strany 35

Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual ix Datalogging Settings ...

Strany 36 - Specifications

Operation Alarms Menu 3-68 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DETECTOR ALARMS: BOARD STATUS OK >ALPHA

Strany 37

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-69 temperature, nephelometer RH, LED current, and reference

Strany 38

Operation Alarms Menu 3-70 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific NEPHELOMETER TEMP: ACTUAL MIN: 0.0 oC SET MIN

Strany 39

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-71 the minimum or maximum limit, an alarm is activated and

Strany 40 - Inspection

Operation Alarms Menu 3-72 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The RH/Temperature Alarms submenu allows the user to vi

Strany 41 - Acceptance Testing

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-73 RELATIVE HUMIDITY: ACTUAL MIN: 5.0 % SET

Strany 42 - Acceptance Test

Operation Alarms Menu 3-74 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific alarm (bell) icon appears in the status bar on the Run s

Strany 43

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-75 The Minimum Flow Temperature alarm limit screen is used

Strany 44

Operation Alarms Menu 3-76 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific BOARD TEMPERATURE: ACTUAL MIN: 5.0 oC SET MIN

Strany 45 - One-Point RH Sensor

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-77 BAROMETRIC PRESSURE: ACTUAL MIN: 400.0 mmHg SET MI

Strany 46 - Flow Rate Verification

Contents x Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Non-Alarm ...

Strany 47

Operation Alarms Menu 3-78 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Flow screen displays the current flow reading and s

Strany 48 - Inlet System

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-79 FLOW ALARMS: >FLOW OK

Strany 49

Operation Alarms Menu 3-80 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CONC ALARMS: >AVG PM OK AVG SHARP

Strany 50

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-81 The Average SHARP screen displays the current average S

Strany 51 - Monitor Installation

Operation Alarms Menu 3-82 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific AVERAGE NEPH CONC: ACTUAL 0.0 >MIN

Strany 52

Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-83 minimum and maximum instant PM concentration screens fun

Strany 53

Operation Alarms Menu 3-84 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INSTANT SHARP CONC: ACT MIN: 0.0 SET MIN

Strany 54

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-85 The Service menu appears only when the instrument is i

Strany 55 - External Devices

Operation Service Menu 3-86 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific RH/TEMP CALIBRATION: >AMBIENT TEMP 4.2 oC FL

Strany 56 - Connecting External Devices

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-87 CALIBRATE AMBIENT RH: REL HUMIDITY: 40.0 %

Strany 57

Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xi Filter Tape Counter ...

Strany 58

Operation Service Menu 3-88 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Calibrate Barometer Pressure Offset screen allows

Strany 59

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-89 ● ● ● ● The Vacuum/Flow Calibration submenu

Strany 60 - Power Connections

Operation Service Menu 3-90 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific under the filter tape from the positive ΔP port on the

Strany 61 - Important Tips

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-91 ● ● ● ● The Flow Calibration submenu is use

Strany 62 - (4–50 °C)

Operation Service Menu 3-92 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Manual Flow Calibration screen allows the user to

Strany 63

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-93 THERMAL MASS COEFFICIENT: CURRENTLY: 0.000000

Strany 64 - Pushbuttons

Operation Service Menu 3-94 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific MASS COEFFICIENT: >AUTO MANUAL

Strany 65 - Soft Keys

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-95 Note This value is set in the factory and should not b

Strany 66 - Overview

Operation Service Menu 3-96 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DETECTOR AUTO CAL: HIGH VOLT: 1500 V BETA CNT :

Strany 67 - Firmware Overview

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-97 HIGH VOLTAGE: CURRENTLY: 1500 V SET TO:

Strany 68 - Run Screens

Contents xii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Min and Max Instant Nephelometer Concentration Limits ... 3-84Se

Strany 69 - 40.0%

Operation Service Menu 3-98 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Service > Nephelometer C

Strany 70

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-99 SOURCE LEVEL SET: IRED: 70 mA SRC LEV

Strany 71 - 22.7 PPM

Operation Service Menu 3-100 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANALOG OUTPUT CAL: ZERO CONNECT METER TO OUTPUT!

Strany 72 - Range Menu

Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-101 ANALOG INPUT CAL: >INPUT CHANNEL 1 INPUT CHANNEL

Strany 73 - SHARP Range

Operation Service Menu 3-102 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Display Pixel Test screen is used to test the LCD

Strany 74 - Set Custom Ranges

Operation Password Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-103 ● ● ● ● ● The Password menu allows the use

Strany 75 - Integration Time

Operation Password Menu 3-104 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Lock Instrument screen is used to lock the instr

Strany 76 - Calibration Factors

Operation Password Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-105 The Remove Password screen is used to erase the curr

Strany 78 - Nephelometer Background

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-1 Chapter 4 Calibration The Model 5030i is an instrument that can make accurate

Strany 79 - ©ª CHANGE VALUE

Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xiii Restore Factory Defaults ...

Strany 80 - Instrument Controls

Calibration Pre-Calibration 4-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific nearest 0.1 °C. This thermometer should be referenc

Strany 81 - Set Flow/Pump

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-3 All screens that appear in this chapter have

Strany 82 - Set Heater

Calibration Calibration Procedure 4-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Assuming the instrument cover has been remov

Strany 83 - Filter Tape Control

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-5 CAL BARO PRESSURE SPAN: PRESSURE: 760

Strany 84

Calibration Calibration Procedure 4-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 4–1. Differential Pressure and Vacuum

Strany 85

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-7 ● In the Main Menu, choose Service > Pres

Strany 86 - Conditions

Calibration Calibration Procedure 4-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Mass Calibration submenu is used to view

Strany 87 - Settings

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-9 Note To achieve the most accurate mass calib

Strany 88

Calibration Calibration Procedure 4-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 3. Cut/break the filter tape and remove fro

Strany 89

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-11 ● ● ● ● The span foil calibration

Strany 90

Contents xiv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Fan Filter Inspection and Cleaning ...

Strany 91

Calibration Calibration Procedure 4-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific MASS COEF: 6878.7 FOIL VALUE: 13

Strany 92

Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-13 The Nephelometer Calibration submenu is use

Strany 93

Calibration Calibration Procedure 4-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Service > Nep

Strany 94

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-1 Chapter 5 Preventive Maintenance This chapter describes the periodic maintena

Strany 95

Preventive Maintenance Replacement Parts 5-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific See the “Servicing” chapter for a lis

Strany 96

Preventive Maintenance Cleaning the Inlets Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-3 Figure 5–1. U.S. EPA PM10 Inlet Onc

Strany 97 - Communication

Preventive Maintenance Cleaning the Inlets 5-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 5–2. European PM10 Inlet PM1

Strany 98

Preventive Maintenance Fan Filter Inspection and Cleaning Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-5 Figure 5–3. PM2.5 Sh

Strany 99

Preventive Maintenance Pump Rebuilding 5-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Equipment Damage Some internal compone

Strany 100

Preventive Maintenance Filter Tape Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-7 3. Record the reference volumet

Strany 101

Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xv Flow RTD Replacement ...

Strany 102 - Volumetric Conditions

Preventive Maintenance Greasing the Cam 5-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 5–5. Replacing the Filter Tap

Strany 103

Preventive Maintenance External Pump Exhaust Filter Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-9 1. Turn the Heater OFF: Fr

Strany 104

Preventive Maintenance Cleaning the SHARP Optics Assembly 5-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following p

Strany 105

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-1 Chapter 6 Troubleshooting This instrument has been designed to achieve a high

Strany 106 - I/O Configuration

Troubleshooting Troubleshooting Guides 6-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 6–1. Troubleshooting - General Gui

Strany 107

Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-3 Malfunction Possible Cause Action LC

Strany 108

Troubleshooting Troubleshooting Guides 6-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Malfunction Possible Cause Action No S

Strany 109

Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-5 Table 6–2. Troubleshooting - Alarm Mess

Strany 110

Troubleshooting Troubleshooting Guides 6-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Alarm Message Possible Cause Action S

Strany 111 - ©ª CHANGE VALUE  SAVE

Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-7 Alarm Message Possible Cause Action Pr

Strany 112

© 2011 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available in

Strany 113

Contents xvi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Analog Voltage Outputs ...

Strany 114

Troubleshooting Board-Level Connection Diagrams 6-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2 PinDI GI T ALOUT P UT S24VDCP

Strany 115

Troubleshooting Board-Level Connection Diagrams Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-9 Figure 6–2. Board-Level Conn

Strany 116

Troubleshooting Connector Pin Descriptions 6-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 6–3. Board-Level Connectio

Strany 117

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-11 Connector Label Reference Designato

Strany 118 - Service Mode

Troubleshooting Connector Pin Descriptions 6-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 119 - Timezone

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-13 Connector Label Reference Designato

Strany 120 - Diagnostics Menu

Troubleshooting Connector Pin Descriptions 6-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 121 - Voltages

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-15 Connector Label Reference Designato

Strany 122

Troubleshooting Connector Pin Descriptions 6-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 123 - Pressure/Vacuum

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-17 Connector Label Reference Designato

Strany 124 - Nephelometer

Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xvii Communications Configuration ...

Strany 125 - Digital Inputs

Troubleshooting Connector Pin Descriptions 6-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 126 - Test Analog Outputs

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-19 Connector Label Reference Designato

Strany 127 - Contact Information

Troubleshooting Connector Pin Descriptions 6-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 6–5. Front Panel Board Connec

Strany 128 - Alarms Menu

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-21 Connector Label Reference Designato

Strany 129 - Detector Alarms

Troubleshooting Connector Pin Descriptions 6-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 130

Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-23 Table 6–7. Digital Output Board Co

Strany 131

Troubleshooting Connector Pin Descriptions 6-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato

Strany 132

Troubleshooting Service Locations Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-25 Connector Label Reference Designator Pin Si

Strany 134

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-1 Chapter 7 Servicing This chapter explains how to replace the Model 5030i suba

Strany 135

Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific

Strany 136

Servicing Service Locations 7-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● “Pressure Board Calibration” on page 7-20 ● “An

Strany 137

Servicing Safety Precautions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-3 Read the safety precautions before beginning any

Strany 138

Servicing Firmware Updates 7-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The LCD polarizing plate is very fragile, handle it

Strany 139

Servicing Replacement Parts List Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-5 Table 7-1 lists the replacement parts for the

Strany 140 - Flow Alarms

Servicing Replacement Parts List 7-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Part Number Description 10-001403 External P

Strany 141 - Concentration Alarms

Servicing Cable List Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-7 Table 7–2 lists the Model 5030i cables. See the “Troubles

Strany 142 - ©ª CHANGE VALUE  SAVE

Servicing External Device Connection Components 7-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 7–3 lists the standard a

Strany 143

Servicing External Device Connection Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-9 Front Panel Board Pressure Boar

Strany 144

Servicing External Device Connection Components 7-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific SHARP Board Bracket AssyRefere

Strany 145

Servicing Removing the SHARP Optics Assembly Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-11 The SHARP optics assembly can be

Strany 146

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xix Figures Figure 1–1. SHARP Monitor Sample Path ...

Strany 147 - Service Menu

Servicing Removing the Measurement Case Assembly and Lowering the Partition Panel 7-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scienti

Strany 148

Servicing Removing the Measurement Case Assembly and Lowering the Partition Panel Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7

Strany 149 - Calibration

Servicing Fuse Replacement 7-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to replace the fuse. E

Strany 150 - Calibrate Barometer

Servicing Detector Amplifier Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-15 5. Remove the four fan moun

Strany 151 - Calibrate Vacuum

Servicing Detector Amplifier Assembly Replacement 7-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Remove the SHARP optics

Strany 152

Servicing Detector Amplifier Board Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-17 Note If Service Mode is not di

Strany 153 - Flow Calibration

Servicing Pressure Board Replacement 7-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Install the new pump by following the

Strany 154 - Mass Calibration

Servicing Pressure Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-19 3. Disconnect the pressure board from th

Strany 155

Servicing Pressure Board Calibration 7-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–10. Replacing the Pressure Boar

Strany 156

Servicing Analog Output Testing Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-21 b. Turn on manometer and allow to self calibr

Strany 157 - Detector Calibration

Figures xx Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–7. Replacing the Detector Amplifier Assembly ...

Strany 158 -  TO START AUTO CAL

Servicing Analog Output Testing 7-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1. Connect a meter to the channel to be teste

Strany 159

Servicing Analog Output Testing Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-23 Figure 7–11. Rear Panel Analog Input and Outp

Strany 160

Servicing Analog Output Calibration 7-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Input Channel Pin 4 5 5 6 6 7 7 9 8 10 Gr

Strany 161 - Analog Output Calibration

Servicing Analog Input Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-25 6. Press to return to the previous scree

Strany 162 - Analog Input Calibration

Servicing Themistor Replacement 7-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to calibrate the

Strany 163

Servicing I/O Expansion Board (Optional) Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-27 1. Turn the instrument O

Strany 164 - Defaults

Servicing I/O Expansion Board (Optional) Replacement 7-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Unplug the I/O expans

Strany 165 - Password Menu

Servicing Digital Output Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-29 Figure 7–14. Rear Panel Board Conn

Strany 166 - Change Password

Servicing Motherboard Replacement 7-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 5. Pop the digital output board off of the

Strany 167 - Unlock Instrument

Servicing Measurement Interface Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-31 Use the following procedure

Strany 168

Figures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xxi Figure 10–6. Radius Tube Adapter Alignment Check ...

Strany 169

Servicing Photo Interrupt Board Replacement 7-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–15. Replacing the Measu

Strany 170 - Procedure

Servicing Proportional Valve Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-33 4. Remove board. 5. Re-install the

Strany 171 - Temperature

Servicing Detector Assembly Replacement 7-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Disconnect plumbing. 5. Remove fo

Strany 172

Servicing Detector Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-35 WARNING The detector assembly should

Strany 173

Servicing Flow RTD Replacement 7-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–18. Replacing the Detector Assembly

Strany 174 - Calibration Procedure

Servicing Transformer Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-37 4. Unplug the RTD connector from the measur

Strany 175

Servicing Radius Tube Replacement 7-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1. Turn the instrument OFF, unplug the powe

Strany 176

Servicing Ambient RH/Temperature Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-39 1. Turn the instrument OFF, unpl

Strany 177 -  TO SAVE FOIL VALUE

Servicing Ambient RH/Temperature Replacement 7-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Equipment Damage Some internal

Strany 178

Servicing Front Panel Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-41 Figure 7–22. Replacing the Ambient RH

Strany 179 - 3μg

Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific

Strany 180

Servicing LCD Module Replacement 7-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–23. Replacing the Front Panel Boar

Strany 181

Servicing LCD Module Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-43 Do not wipe the LCD polarizing plate with a d

Strany 182

Servicing Removing the SHARP Optics Cover 7-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP optics cover can be remov

Strany 183 - Preventive Maintenance

Servicing External Solenoid Valve Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-45 Use the following procedure to

Strany 184 - Cleaning the Inlets

Servicing SHARP RH/Temp Sensor Replacement 7-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to rep

Strany 185 - European PM

Servicing SHARP Interface Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-47 Figure 7–26. Disconnecting the R

Strany 186 - Maintenance

Servicing SHARP Interface Board Replacement 7-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Unplug all connectors from the

Strany 187 - Fan Filter Inspection

Servicing Delrin Nut Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-49 Figure 7–28. Replacing the SHARP Optics Bo

Strany 188 - Leak Test

Servicing Delrin Nut Replacement 7-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Turn the Heater OFF: From the Main Menu,

Strany 189 - Filter Tape

Servicing Internal SHARP Cable Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-51 Use the following procedu

Strany 190 - Exhaust Filter

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xxiii Tables Table 1–1. Model 5030i Specifications ...

Strany 191 - HEPA Filter

Servicing Case Cable Replacement 7-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–30. Replacing the Internal SHARP Ca

Strany 192 - Assembly

Servicing External Zeroing Filter Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-53 4. Replace the case cable assembly by follo

Strany 193 - Troubleshooting

Servicing Service Locations 7-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–32. Replacing the External Zeroing Filte

Strany 194 - Troubleshooting Guides

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-1 Chapter 8 System Description This chapter describes the function and location

Strany 195

System Description Hardware 8-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Front Panel Board Pressure BoardFront Panel Tube Mo

Strany 196

System Description Hardware Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-3 SHARP Board Bracket Assy Reference Detector BoardSH

Strany 197

System Description Hardware 8-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The nephelometer assembly uses an 880 nanometer in

Strany 198

System Description Firmware Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-5 This PCB assembly is mounted on the floor plate an

Strany 199

System Description Electronics 8-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Signals are gathered from the low-level process

Strany 200 - TO MEASUREMENT SYSTEM

System Description Electronics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-7 External connectors include: ● External Accesso

Strany 201

Tables xxiv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table B–8. Reply Termination Formats ...

Strany 202 - Descriptions

System Description Electronics 8-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connectors located on the measurement interface

Strany 203 - Connector Pin Descriptions

System Description I/O Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-9 The I/O expansion board connects to the moth

Strany 204

System Description I/O Components 8-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The analog outputs may be assigned to any me

Strany 205

System Description I/O Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-11 Sixteen digital inputs are available, which

Strany 206

System Description I/O Components 8-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 7 or 8 data bits 1 or 2 stop bits No, odd, o

Strany 207

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-1 Chapter 9 Optional Equipment The Model 5030i is available with the following

Strany 208

Optional Equipment Terminal Block and Cable Kits 9-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific connector board. For associate

Strany 209

Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-3 Table 9–2. Color Codes for 25-Pin and 37-P

Strany 210

Optional Equipment Mounting Options 9-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 9–1. Bench Mounting

Strany 211

Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-5 Figure 9–2. EIA Rack Mounting

Strany 212

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-1 Chapter 1 Introduction The Model 5030i Synchronized Hybrid Ambient Real-time

Strany 213

Optional Equipment Mounting Options 9-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 9–3. Retrofit Rack Mounting

Strany 214

Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-7 Figure 9–4. Rack Mount Option Assembly

Strany 215

Optional Equipment Roof Flange Assembly 9-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific An optional roof flange assembly can b

Strany 216

Optional Equipment Roof Flange Assembly Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-9 The flange design also has four holes t

Strany 218

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-1 Chapter 10 Model 5014i to 5030i Upgrade Procedures This chapter describes th

Strany 219 - Servicing

Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation 10-2 Model 5030i SHARP Instruction Manual Thermo Fisher Sc

Strany 220 - Service Locations

Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual

Strany 221 - Safety Precautions

Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation 10-4 Model 5030i SHARP Instruction Manual Thermo Fisher Sc

Strany 222 - Accessing the

Model 5014i to 5030i Upgrade Procedures Radius Tube Adapter Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-5 CONN

Strany 223 - Replacement Parts

Thermo Fisher Scientific WEEE Compliance WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electro

Strany 224 - Replacement Parts List

Introduction Principle of Operation 1-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Diaphragm Vacuum Pump ● External Ambien

Strany 225 - Cable List

Model 5014i to 5030i Upgrade Procedures Radius Tube Adapter Installation 10-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1 =

Strany 226 - Components

Model 5014i to 5030i Upgrade Procedures SHARP External Valve Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-7 Use

Strany 227

Model 5014i to 5030i Upgrade Procedures SHARP External Valve Installation 10-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific c.

Strany 228

Model 5014i to 5030i Upgrade Procedures SHARP Optics Assembly Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-9 Us

Strany 229 - Optics Assembly

Model 5014i to 5030i Upgrade Procedures Upgrade of the Heater Tube Assembly 10-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific

Strany 230 - Assembly and Lowering

Model 5014i to 5030i Upgrade Procedures Model Label Change Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-11 Figure 10–12. Con

Strany 232 - Fan/Filter

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual A-1 Appendix A Warranty Seller warrants that the Products will operate or perform

Strany 233 - Detector Amplifier

Warranty Warranty A-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific they were not designed, (v) causes external to the Products s

Strany 234

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-1 Appendix B C-Link Protocol Commands This appendix provides a description of t

Strany 235 - Board Calibration

Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-3 For details of the analyzer’s principle of

Strany 236 - Pressure Board

C-Link Protocol Commands Commands B-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific For example, if the instrument ID is set to 2

Strany 237 - Remove Screws (4)

C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-3 Table B–1. Error Response Messages Command Re

Strany 238

C-Link Protocol Commands Commands B-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page alarm amb rh min

Strany 239 - Analog Output Testing

C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-5 Command Description Page alarm filter tape

Strany 240

C-Link Protocol Commands Commands B-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page alpha eff Reports

Strany 241

C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-7 Command Description Page copy lrec to sp S

Strany 242 - Analog Output

C-Link Protocol Commands Commands B-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page erec layout Repor

Strany 243

C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-9 Command Description Page list var log Repo

Strany 244 - Themistor

C-Link Protocol Commands Commands B-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page no or srec Repor

Strany 245 - (Optional)

C-Link Protocol Commands Measurements Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-11 Command Description Page tape counter

Strany 246 - I/O Expansion Board (Opt)

Introduction Principle of Operation 1-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP optical assembly senses the light

Strany 247

C-Link Protocol Commands Measurements B-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific flags This reports 8 hexadecimal digits

Strany 248

C-Link Protocol Commands Measurements Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-13 Send: neph 24avg Receive: neph 24avg

Strany 249 - Interface Board

C-Link Protocol Commands Alarms B-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: pm 24avg Receive: pm 24avg 10.000E+00

Strany 250 - Board Replacement

C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-15 alarm amb temp min alarm amb temp max These co

Strany 251 - Proportional Valve

C-Link Protocol Commands Alarms B-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific These commands report the current beta count a

Strany 252 - Detector Assembly

C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-17 alarm conc avg pm min alarm conc avg pm max Th

Strany 253

C-Link Protocol Commands Alarms B-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: alarm conc inst neph min Receive: al

Strany 254 - Flow RTD

C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-19 representation of the concentration alarm limi

Strany 255 - Transformer

C-Link Protocol Commands Alarms B-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set alarm flow pres max 50 Receive:

Strany 256 - Radius Tube

C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-21 Send: alarm nb status Receive: alarm nb st

Strany 257 - Ambient

Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-5 HEPA FILTERβ SOURCE Figure 1–2. Nephelomet

Strany 258

C-Link Protocol Commands Alarms B-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific alarm neph temp min alarm neph temp max This c

Strany 259 - Front Panel Board

C-Link Protocol Commands Diagnostics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-23 alarm tape counter max This command repor

Strany 260 - LCD Module

C-Link Protocol Commands Diagnostics B-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific baro pres This command reports the curren

Strany 261

C-Link Protocol Commands Diagnostics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-25 Send: diag volt mb Receive: diag volt

Strany 262 - Optics Cover

C-Link Protocol Commands Diagnostics B-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: io bd Receive: io bd yes mass T

Strany 263 - Solenoid Valve

C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-27 tape counter This command reports the num

Strany 264 - SHARP RH/Temp

C-Link Protocol Commands Datalogging B-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set copy lrec to sp set copy srec to sp s

Strany 265 - SHARP Interface

C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-29 The format is defined within the command

Strany 266

C-Link Protocol Commands Datalogging B-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific bb = minutes (01 to 59) oo = month (01 to

Strany 267 - Delrin Nut

C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-31 set lrec format format set srec format fo

Strany 268 - Delrin Nut Replacement

Introduction Principle of Operation 1-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific A clean filter spot is introduced by an aut

Strany 269 - Cable Assembly

C-Link Protocol Commands Datalogging B-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific These commands report the lrecs and srecs

Strany 270 - Case Cable

C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-33 sp field number This command reports the

Strany 271 - External Zeroing

C-Link Protocol Commands Calibration B-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific stream time This command reports if the s

Strany 272

C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-35 set alpha th value This command sets the

Strany 273 - System Description

C-Link Protocol Commands Calibration B-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set baro mass coef value This command set

Strany 274 - Hardware

C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-37 set bkg pm value This command is used to

Strany 275

C-Link Protocol Commands Calibration B-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set cal flow span 17.2 Receive:

Strany 276

C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-39 flow pres span This command reports the

Strany 277 - Firmware

C-Link Protocol Commands Calibration B-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set flow temp offset value This command s

Strany 278 - Electronics

C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-41 Send: set thermal mass coef 20.5 Recei

Strany 279 - Measurement Interface

Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-7 events, a rear-panel mounted zeroing soleno

Strany 280

C-Link Protocol Commands Calibration B-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific neph restore bkg This command will restor

Strany 281 - I/O Components

C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-43 set neph temp offset value This command s

Strany 282

C-Link Protocol Commands Keys/Display B-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set vac pres span value This command set

Strany 283 - RS-232 Connection

C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-45 form to save time in transm

Strany 284 - Connector

C-Link Protocol Commands Measurement Configuration B-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: conc unit Receive:

Strany 285 - Optional Equipment

C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-47 filter time This command r

Strany 286 - Cable Kits

C-Link Protocol Commands Measurement Configuration B-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific This command sets the heate

Strany 287 - Mounting Options

C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-49 These commands turn the pre

Strany 288 - Figure 9–1. Bench Mounting

C-Link Protocol Commands Measurement Configuration B-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Selection μg/m3 mg/m3 6 C

Strany 289

C-Link Protocol Commands Hardware Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-51 set temp std value This comman

Strany 290

Introduction Specifications 1-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The beta measurement chamber houses an RTD temperat

Strany 291

C-Link Protocol Commands Hardware Configuration B-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Level Contrast Level 5 25% 6

Strany 292 - Roof Flange

C-Link Protocol Commands Hardware Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-53 Send: set default params Re

Strany 293 - Roof Flange Assembly

C-Link Protocol Commands Communications Configuration B-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific This command sets the in

Strany 294

C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-55 Note This command canno

Strany 295 - Procedures

C-Link Protocol Commands Communications Configuration B-56 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set allow mode cmd valu

Strany 296 - Installation

C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-57 This command sets the in

Strany 297

C-Link Protocol Commands Communications Configuration B-58 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set format format This c

Strany 298

C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-59 set instrument id value

Strany 299 - Adapter Installation

C-Link Protocol Commands Communications Configuration B-60 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set power up mode value

Strany 300

C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-61 analog iout range channel This com

Strany 301 - Valve Installation

Introduction Specifications Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-9 Connecting and Interface Cables, and Vacuum Pump 3-

Strany 302

C-Link Protocol Commands I/O Configuration B-62 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: analog vout range 2 Receiv

Strany 303

C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-63 dout channel This command reports

Strany 304 - Heater Tube

C-Link Protocol Commands I/O Configuration B-64 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific D to A Function Range 8 Current

Strany 305 - Instrument

C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-65 17 frol 19 beta 28 braw 29

Strany 306

C-Link Protocol Commands Record Layout Definition B-66 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set relay open 1 Re

Strany 307 - Warranty

C-Link Protocol Commands Record Layout Definition Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-67 E - parse a 24-bit floatin

Strany 308

C-Link Protocol Commands Record Layout Definition B-68 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Typically, bitfield extracti

Strany 309 - C-Link Protocol Commands

C-Link Protocol Commands Record Layout Definition Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-69 I—Indicates a button which p

Strany 310 - Commands

C-Link Protocol Commands Record Layout Definition B-70 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ' NO:21f{Code_0 Code

Strany 311 - Commands List

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-1 Appendix C MODBUS Protocol This appendix provides a description of the MODBUS

Strany 312

Introduction Specifications 1-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Filter Media Sample Filter: Glass fiber filter t

Strany 313

MODBUS Protocol TCP Communication Parameters C-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific iSeries instruments support the M

Strany 314

MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-3 The function code is a single byte in length. T

Strany 315

MODBUS Protocol Function Codes C-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific outputs follow toward the high end of this byte,

Strany 316

MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-5 Here is an example of a request and response to

Strany 317

MODBUS Protocol Function Codes C-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Request Function Code 1 Byte 0x03 or 0x04 S

Strany 318

MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-7 Register Value Hi (12) 0x00 Register Value Lo (

Strany 319 - Measurements

MODBUS Protocol MODBUS Parameters Supported C-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Here is an example of a request to

Strany 320

MODBUS Protocol MODBUS Parameters Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-9 Coil Number Status 7 BETA DETECTO

Strany 321

MODBUS Protocol MODBUS Parameters Supported C-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Register Number Variable 23 &

Strany 322

MODBUS Protocol MODBUS Parameters Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-11 Register Number Variable 91 &

Strany 323

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-1 Chapter 2 Installation The following installation procedures for the Model 50

Strany 324

MODBUS Protocol MODBUS Parameters Supported C-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Coil Number Action Triggered 106

Strany 325

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-1 Appendix D Gesytec (Bayern-Hessen) Protocol This appendix provides a descript

Strany 326

Gesytec (Bayern-Hessen) Protocol TCP Communication Parameters D-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific iSeries instrume

Strany 327

Gesytec (Bayern-Hessen) Protocol Block Checksum Characters <BCC> Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-3 The Blo

Strany 328

Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific If the received command does

Strany 329

Gesytec (Bayern-Hessen) Protocol Gesytec Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-5 Sample Data Reply String in r

Strany 330

Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Gesytec serial number def

Strany 331 - Diagnostics

Gesytec (Bayern-Hessen) Protocol Gesytec Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-7 The following 14 measurement

Strany 332

Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific See Table D–1 for operating

Strany 333

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual E-1 Appendix E ESM Protocol Commands This appendix provides a description of the

Strany 335 - Datalogging

Installation Lifting 2-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The inner box that the instrument resides in is an ISTA 1A

Strany 336

ESM Protocol Commands ESM Commands Supported E-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description UA beta coun

Strany 337

ESM Protocol Commands ESM Commands Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual E-3 Command Description E pump on F f

Strany 340

Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-3 Remove Packing(2 pieces)

Strany 341

Installation Acceptance Testing and Startup Procedures 2-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific program. Furthermore, it

Strany 342

Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-5 Note If the Model 5030i

Strany 343

Installation Acceptance Testing and Startup Procedures 2-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Press to return to

Strany 344

Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-7 Figure 2–2. Model 5030i

Strany 345

Installation Acceptance Testing and Startup Procedures 2-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific If the RH sensors are sl

Strany 346

Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-9 Attach the small sample

Strany 347

Installation Setup 2-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific nephelometer zero should be conducted. After re-zeroing the

Strany 348

Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-11 Figure 2–4. Model 5030i Top View—inches [millimeters] Once

Strany 349

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual i About This Manual This manual provides information about installing, operating,

Strany 350

Installation Setup 2-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–5. Model 5030i Vertical View Ambient Inlet (PM1

Strany 351

Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-13 There is one standard length of heated sample tube at one

Strany 352 - Keys/Display

Installation Setup 2-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–6. Heater Assembly 4. Attach a second tubing con

Strany 353 - Send: screen

Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-15 ambient T/RH sensor assembly. Tighten this adapter against

Strany 354 - Measurement Configuration

Installation Setup 2-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–7. Rear Panel Plumbing Schematic Heater Assemb

Strany 355

Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-17 Several components are available for

Strany 356

Installation Connecting External Devices 2-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–8. I/O Termin

Strany 357

Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-19 The analog outputs are arranged on th

Strany 358

Installation Connecting External Devices 2-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–10. D/O Termi

Strany 359 - Configuration

Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-21 Assembled Connector down and spacer.

Strany 360 - Hardware Configuration

About This Manual Safety ii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Appendix B “C-Link Protocol Commands” provides a des

Strany 361

Installation Startup 2-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Before the power is connected, be sure that the front pa

Strany 362 - Communications

Installation Data Content Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-23 set to 60 minute logging with 30% memory and Srecs a

Strany 363 - Communications Configuration

Installation Important Tips 2-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DO ● Operate the instrument in a climate-controll

Strany 364

Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-1 Chapter 3 Operation This chapter describes the front panel display screens, f

Strany 365

Operation Pushbuttons 3-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 3–1. Front Panel Display CAUTION If the LCD pan

Strany 366

Operation Pushbuttons Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-3 Table 3–1 lists the front panel pushbuttons and their fun

Strany 367

Operation Firmware Overview 3-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANGE abcdefghijklmn BKSP opq

Strany 368

Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-5 submenu contains related instrument settings. This

Strany 369

Operation Firmware Overview 3-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Power-Up screen is displayed when power is a

Strany 370

Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-7 The Run screen normally displays the current SH

Strany 371

About This Manual Safety and Equipment Damage Alerts Thermo Fisher Scientific Model 5030i SHARP Instruction Manual iii Alert Description If

Strany 372

Operation Firmware Overview 3-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Ambient Conditions Run screen displays the

Strany 373

Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-9 The Hybrid Data Run screen displays the mass con

Strany 374 - Definition

Operation Range Menu 3-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific XXLO WCONCENTRATIONXXLO W SHARP 15.0 3µg

Strany 375 - Erec Layout

Operation Range Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-11 CONC UNITS: CURRENTLY: mg/m3 SET TO:

Strany 376 - Record Layout Definition

Operation Range Menu 3-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The analog outputs are arranged on the rear panel connect

Strany 377 - ' NO:3s\n'

Operation Integration Time Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-13 ● In the Main Menu, choose Range > Set Custom R

Strany 378

Operation 24-Hour Average 3-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The 24-Hour Average screen defines a start time dur

Strany 379 - MODBUS Protocol

Operation Calibration Factors Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-15 CALIBRATION FACTORS: >PM BKG

Strany 380 - Unit Definition

Operation Calibration Factors Menu 3-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP Background screen allows the use

Strany 381 - Function Codes

Operation Calibration Factors Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-17 NEPHELOMETER BKG: >NEPH MODE VALUES R

Strany 382

About This Manual Electrical/Safety Certifications iv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Alert Description use a so

Strany 383 - Registers

Operation Instrument Controls Menu 3-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific RESTORE PREV BKG VALUES?

Strany 384 - *N = Quantity of Registers

Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-19 INSTRUMENT CONTROLS: >SET FLOW/PUMP SET

Strany 385 - Single Coil

Operation Instrument Controls Menu 3-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific PUMP: CURRENTLY: ON SET TO:

Strany 386 - Supported

Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-21 RH THRESHOLD: CURRENTLY: 65 %

Strany 387 - MODBUS Parameters Supported

Operation Instrument Controls Menu 3-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific START DATE AND TIME: 01 NOV 2008 00:00 ?

Strany 388

Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-23 ● In the Main Menu, choose Instrument Cont

Strany 389

Operation Volumetric Conditions 3-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Volumetric Conditions menu allows the use

Strany 390

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-25 The Pressure Compensation screen is used to t

Strany 391 - Serial Communication

Operation Volumetric Conditions 3-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific containing 32 items once each minute for a wee

Strany 392 - TCP Communication Parameters

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-27 SELECT START POINT BY: SET TO: # OF RE

Strany 393 - Gesytec Commands

About This Manual WEEE Symbol Thermo Fisher Scientific Model 5030i SHARP Instruction Manual v The following symbol and description identify the

Strany 394 - Query Command

Operation Volumetric Conditions 3-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ERASE LREC LOG FILE DATA?

Strany 395

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-29 LREC FIELDS: >FIELD 1 SHARP FIEL

Strany 396

Operation Volumetric Conditions 3-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Other Measurements screen allows the user

Strany 397 - Measurements reported in

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-31 Figure 3–5. Common Flags Figure 3–6. Detect

Strany 398 - Operating and Error

Operation Volumetric Conditions 3-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific by “<--” after it. Note that at this point,

Strany 399 - ESM Protocol Commands

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-33 ● ● ● ● The Configure Datalogging m

Strany 400 - ESM Commands Supported

Operation Volumetric Conditions 3-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Control

Strany 401 - F filter change

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-35 SET LREC FLAG STATUS DATA: CURRENTLY: OFF

Strany 402

Operation Volumetric Conditions 3-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific BAUD RATE: CURRENTLY: 9600

Strany 403

Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-37 STOP BITS: CURRENTLY: 1