Model 5014i to 5030i Upgrade Procedures
SHARP Connection Bracket Assembly Installation
10-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
6. Unplug instrument from AC power source.
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
DETECTOR BETA
DETECTOR B NONE
§¨ CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Use the following procedure for SHARP connection bracket assembly
installation.
1. Locate the Connection Bracket (Figure 10–1) and remove two screws
and set aside.
4-pin SHARP
Connector
2-pin connector
not used
Figure 10–1. SHARP Optics Connection Bracket
2. Remove both sides of the Cover. Remove the tape pick up spool
(Figure 10–2) by removing the thumb screw, washer, filter tape,
spacers, o-rings and rear filter support guide plate.
SHARP Connection
Bracket Assembly
Installation