Sharp SYSTEM 8/10 Plus Manuál s instrukcemi Strana 364

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 403
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 363
C-Link Protocol Commands
Communications Configuration
B-56 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
set allow mode cmd value
This command is used to configure the instrument to value, where value is
either 1 = accept or 0 = ignore the “set mode local” and “set mode remote”
commands, according to Table B–7.
If the instrument is set to accept the commands (value = 1), the “set mode
local” command will unlock the instrument and the keypad can be used to
make changes via the front panel.
If the instrument is set to ignore the commands (value = 0), the instrument
will respond with “ok” as if the command has been accepted and acted
upon, but will not change the instrument lock status (this is for
compatibility with systems expecting and “ok” response).
Note The instrument will always respond to the command “mode” with
the status of the password lock as “mode local” or “mode remote”,
regardless of the above setting.
The following example sets the instrument to accept the “set mode local”
and “set mode remote” commands.
Send: set allow mode cmd 1
Receive: set allow mode cmd 1 ok
Table B–7. Allow Mode Command Values
Value
Allow Mode Command
0 Ignore (default)
1 Accept
baud
This command reports the current baud rate for the serial port
(RS232/RS485). The following example reports that the current baud rate
is 9600 baud.
Send: baud
Receive: baud 9600
set baud rate
rate = | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
Zobrazit stránku 363
1 2 ... 359 360 361 362 363 364 365 366 367 368 369 ... 402 403

Komentáře k této Příručce

Žádné komentáře