CarboVis 70x IQ TS What to do if...
7 - 3
ba75950e02 11/2012
Measured values too low
or too high
Display of
ERROR
Cause Remedy
Gas bubbles in front of the
measurement windows
Check the installation position
of the sensor
Measurement windows dirty Clean the measurement win-
dows
Cause Remedy
UV-VIS sensor not correctly
connected
Check the assignment of the
terminal connections
Load at a power supply mod-
ule too high
Install another power supply
module in the vicinity of the
sensor
–IQS
ENSOR NET cable to-
wards the CarboVis 70x IQ
TS too long (voltage drop too
great)
Install another power supply
module in the vicinity of the
sensor
Electrical connection be-
tween power supply module
and CarboVis 70x IQ TS is in-
terrupted
Check the cable connection
step-by-step starting with the
power supply module, and
replace any defective cable
sections.
Check the contacts on the MIQ
modules (stack mounting).
Clean any dirty contacts. Care-
fully bend back contact springs
that have been pressed flat or
bent (pay attention to sufficient
spring tension)
Zobrazit stránku 46