CarboVis 70x IQ TS What to do if...
7 - 1
ba75950e02 11/2012
7 What to do if...
"----" display
(invalid measured value)
Implausible measured
values
Cause Remedy
User calibration values
entered incorrectly
*
Sensor check (H2O dest.)
or
*
Zero adjustment
or
*
Sensor check (standard)
defective
Correct the entry and enter
once again (section 4.3.2)
Repeat
*
Sensor check (H2O dest.)
or
*
Zero adjustment
or
*
Sensor check (standard)
.
Check conditions, especially
cleanliness (section 4.3.3)
Zero adjustment
still errone-
ous after several trials
Contact WTW
Unknown See log book
Cause Remedy
Erroneous user calibration Repeat user calibration. Make
sure the samples for compari-
son are representative.
Substantial change of test
sample (sample matrix)
Carry out new user calibration
Measurement windows dirty Clean the measurement win-
dows
Sensor setting wrong Check and, if necessary, cor-
rect the
Measuring mode
,
Mea
-
suring location
sensor settings
Carry out new user calibration
Optical system maladjusted,
e. g. due to inadmissible
mechanical shock
Carry out
*
Sensor check (H2O dest.)
or
*
Zero adjustment
or
*
Sensor check (standard)