Sharp IQ-8000 Servisní příručka Strana 42

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 66
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 41
Maintenance and cleaning CarboVis 70x IQ TS
5 - 4
ba75950e02 11/2012
5.2.4 Cleaning the measuring gap
To remove lime or grease deposits, use the flocked cleaning cards
together with the following detergents:
Proceeding Start by removing any lime deposits.
Soak a flocked cleaning card with the relevant detergent and use it
to clean the measuring windows according to step 6. in section
5.2.3 B
ASIC CLEANING. If necessary, leave the cleaning card in the
measuring gap for a few minutes so the detergent can take effect.
Thoroughly rinse with ultrapure water the measuring gap and the
surrounding areas moistened with detergent.
Then wipe the measurement window with a lens tissue. The surface
of the measurement windows must be free of streaks.
Protect the cleaned sensor against new contamination.
Contamination Cleaning agents
Lime deposits Hydrochloric acid 5 % for analysis
Grease deposits Mixture of isopropanol and water (approx.
70 % isopropanol), made of 7 volume parts
isopropanol for analysis and 3 volume parts
ultrapure water
Zobrazit stránku 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 65 66

Komentáře k této Příručce

Žádné komentáře